Siemens Healthcare Diagnostics Inc. has confirmed the potential for erroneous results due to discolored reagent observed in some wells 5 and 6 of Dimension Vista High Density Lipoprotein…
Class IIFDA: OngoingSiemens Healthcare Diagnostics, Inc.
Gamma-Glutamyl Transferase (GGT) Reagent Low End Repeatability-Imprecision. Reagent may demonstrate a coefficient of variation (%CV) outside of the performance data for repeatability…
Class IIFDA: OngoingSiemens Healthcare Diagnostics, Inc.
Gamma-Glutamyl Transferase (GGT) Reagent Low End Repeatability-Imprecision. Reagent may demonstrate a coefficient of variation (%CV) outside of the performance data for repeatability…
Class IIFDA: OngoingSiemens Healthcare Diagnostics, Inc.
Test Definition scanning may reset custom settings to defaults causing falsely low or high results. When a 2D Master Curve and TDef barcode for a new kit lot of reagent is scanned and the…
Class IIFDA: TerminatedSiemens Healthcare Diagnostics, Inc.
Test Definition scanning may reset custom settings to defaults. When a 2D Master Curve and TDef barcode for a new kit lot of reagent is scanned and the TDef version is a newer version than…
Class IIFDA: TerminatedSiemens Healthcare Diagnostics, Inc.
Low outlier resultsgenerated with Alkaline Phosphatase (ALPI) Flex reagent cartridge.The maximum difference observed with a patient sample was approximately -27% at a concentration of 129…
Class IIFDA: TerminatedSiemens Healthcare Diagnostics, Inc.
Siemens confirmed that the Upper Limit of Normal (ULN) as claimed in the Instructions for Use (IFU) was no longer achieved. This can potentially result in false positive or false negative…
Class IIFDA: TerminatedSiemens Healthcare Diagnostics, Inc.
Fructosamine QC and patient samples may exhibit a 60 umol/L positive bias across the measuring range compared to predicate assay reference leading to a potential adjustment of hyperglycemic…
Class IIFDA: OngoingSiemens Healthcare Diagnostics, Inc.
Fructosamine QC and patient samples may exhibit a 60 umol/L positive bias across the measuring range compared to predicate assay reference leading to a potential adjustment of hyperglycemic…
Class IIFDA: OngoingSiemens Healthcare Diagnostics, Inc.
Positive bias across the Analytical Measurement Range of the TNIH assay compared to an unaffected control lot. The average bias observed for patient samples using lots EB0255 and EA0178 of…
Class IIFDA: TerminatedSiemens Healthcare Diagnostics, Inc.
Sporadically inconsistent discrepant (low bias) glucose results on card lot 01-20095-10. Potential exists for a delay in diagnosis of hyperglycemia or unnecessary treatment for hypoglycemia.
Class IIFDA: TerminatedSiemens Healthcare Diagnostics Inc.
If DCA Vantage Analyzer is configured to run Code 39 with check digit, the scanner is not confirming the integrity of read barcode data using the check digit. Also, if DCA Vantage Analyzer…
Class IIFDA: TerminatedSiemens Healthcare Diagnostics, Inc.
Unexpected interface driver behavior identified (QC) Results May Be Assigned to an Incorrect Control Lot Number, and lead to the reporting of erroneous patient results if the QC failed but…
Class IIFDA: TerminatedSiemens Healthcare Diagnostics, Inc.
Flexible tubing connected to the liquid waste bottle can crack during routine customer maintenance potentially causing liquid waste can leak onto the floor, creating a slip and fall hazard
Class IIFDA: TerminatedSiemens Healthcare Diagnostics Inc.
Unexpected interface driver behavior identified (QC) Results May Be Assigned to an Incorrect Control Lot Number, and lead to the reporting of erroneous patient results if the QC failed but…
Class IIFDA: TerminatedSiemens Healthcare Diagnostics, Inc.
Flexible tubing connected to the liquid waste bottle can crack during routine customer maintenance potentially causing liquid waste can leak onto the floor, creating a slip and fall hazard
Class IIFDA: TerminatedSiemens Healthcare Diagnostics Inc.
Insufficient sample may not be detected when using Tube Top Sample Cup (TTSC) or V-shaped Sample Cup sample containers with the Atellica IM1300 and IM1600 analyzers.
Class IIFDA: OngoingSiemens Healthcare Diagnostics, Inc.
Insufficient sample may not be detected when using Tube Top Sample Cup (TTSC) or V-shaped Sample Cup sample containers with the Atellica IM1300 and Atellica IM1600 analyzers.
Class IIFDA: OngoingSiemens Healthcare Diagnostics, Inc.
The presence of N-acetyl-p-benzoquinone imine (NAPQI), a metabolite of acetaminophen, may cause interference in certain chemistry assays when testing patient samples.
Class IIFDA: TerminatedSiemens Healthcare Diagnostics, Inc.
The presence of N-acetyl-p-benzoquinone imine (NAPQI), a metabolite of acetaminophen, may cause interference in certain chemistry assays when testing patient samples.
Class IIFDA: TerminatedSiemens Healthcare Diagnostics, Inc.
Understanding this recall history
This page summarizes recall records in which the official source identifies Siemens Healthcare Diagnostics, Inc. as the recalling firm or manufacturer. The number of records should not be treated as a safety rating or a measure of company size, because recall counts depend on product mix, reporting periods and how government records are organized.
Compare the dates, categories and classifications together rather than relying on the total alone. Open an individual record for the recalled product, reason, distribution details and current instructions available from the source.