Recall record at a glance
| Official record | Z-0052-2021 |
|---|---|
| Category | Medical Device Recalls |
| Classification | Class II |
| FDA record status | Terminated |
| Recalling firm | Siemens Healthcare Diagnostics, Inc. |
| Report date | Oct 14, 2020 |
| Recall initiation date | Aug 19, 2020 |
| Quantity | 533 US and 1822 OUS |
| Location | Newark, DE, United States |
Official product description
Dimension LOCI High Sensitivity Troponin I (TNIH) Flex reagent cartridge-IVD-In vitro diagnostic use in the quantitative measurement of cardiac troponin I in human plasma Catalog Number: RF627, SMN# 10471068
Why the product was recalled
Positive bias across the Analytical Measurement Range of the TNIH assay compared to an unaffected control lot. The average bias observed for patient samples using lots EB0255 and EA0178 of the Dimension EXL Troponin compared with a control lot was +25%. A maximum bias of +34% in patient samples around the 99th percentile observed
Distribution information
US Nationwide distribution.
How to read the classification
FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.
Additional record details
Recall type: Voluntary: Firm initiated. Initial notification: Letter.
What to do next
If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.
About this record
Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.
FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.