Recall record at a glance
| Official record | Z-0102-2021 |
|---|---|
| Category | Medical Device Recalls |
| Classification | Class II |
| FDA record status | Ongoing |
| Recalling firm | Siemens Healthcare Diagnostics, Inc. |
| Report date | Oct 21, 2020 |
| Recall initiation date | Sep 22, 2020 |
| Quantity | US: 28; OUS: 255 units |
| Location | Tarrytown, NY, United States |
Official product description
Atellica CH Systems Fructosamine (FRUC)-For in vitro diagnostic use in the quantitative measurement of glycated protein (fructosamine) in human serum or plasma on the ADVIA Chemistry systems Siemens Material Number (SMN): 11097637
Why the product was recalled
Fructosamine QC and patient samples may exhibit a 60 umol/L positive bias across the measuring range compared to predicate assay reference leading to a potential adjustment of hyperglycemic therapy. Positively biased fructosamine results may prompt changes in glucose management that may affect glycemia. In extreme situations, more aggressive management of glycemia may occur and contribute to hypoglycemia.
Distribution information
CA, DE, MD, NC, NJ, NY, PA Foreign: Argentina¿¿¿¿¿¿ Australia¿¿¿¿¿¿ Brazil¿¿¿¿ Canada¿¿¿¿¿ Czech¿Republic Denmark¿¿¿¿¿¿¿¿ Germany¿¿¿¿¿¿¿¿ Italy¿¿¿¿¿¿¿¿¿¿ Netherlands¿¿¿¿ P.R.¿China¿¿¿¿¿ Paraguay¿¿¿¿¿¿¿ Slovakia¿¿¿¿¿¿¿ Turkey¿¿¿¿¿¿¿¿¿ United¿Kingdom Vietnam¿¿¿¿¿¿¿¿
How to read the classification
FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.
Additional record details
Recall type: Voluntary: Firm initiated. Initial notification: Letter.
What to do next
If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.
About this record
Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.
FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.