FDA recall record

Siemens Dimension Vista Akaline Phosphatase (ALPI) Flex reagent cartridge - In vitro recall — FDA Z-0134-2021

Low outlier resultsgenerated with Alkaline Phosphatase (ALPI) Flex reagent cartridge.The maximum difference observed with a patient sample was…

Class IIclassification
Z-0134-2021official record ID
Oct 28, 2020report date

Recall record at a glance

Official recordZ-0134-2021
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusTerminated
Recalling firmSiemens Healthcare Diagnostics, Inc.
Report dateOct 28, 2020
Recall initiation dateAug 25, 2020
QuantityUS:5970 units ; OUS: 1446 units
LocationNewark, DE, United States

Official product description

Siemens Dimension Vista Akaline Phosphatase (ALPI) Flex reagent cartridge - In vitro diagnostic test for the quantitative measurement of alkaline phosphatase in human serum and plasma Catalog Number: K2115, SMN# 10642444

Why the product was recalled

Low outlier resultsgenerated with Alkaline Phosphatase (ALPI) Flex reagent cartridge.The maximum difference observed with a patient sample was approximately -27% at a concentration of 129 U/L (2.15 ukat/L). The low outlier results are generated from the last five (5) tests of the well

Distribution information

Worldwide distribution - US Nationwide and the countries of Canada, Australia, Austria, Belgium, Bahamas, Czech Republic, Denmark, France, Germany, Italy, Japan, Korea, Netherlands, Norway, New Zealand, Portugal, Slovakia, Slovenia, Spain, Switzerland.

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.