Siemens Healthcare Diagnostics Atellica IM Enhanced Estradiol (eE2) (REF) 500T - In vitro recall — FDA Z-1022-2022
Falsely Elevated Results Observed with Plasma Specimens, could lead a clinician to misinterpret patient results
Overview of recall records associated with Siemens Healthcare Diagnostics, Inc., including dates, categories and classifications where available.
Classification mix: Class II: 324, Class III: 2.
Falsely Elevated Results Observed with Plasma Specimens, could lead a clinician to misinterpret patient results
Falsely Elevated Results Observed with Plasma Specimens, could lead a clinician to misinterpret patient results
Falsely Elevated Results Observed with Plasma Specimens, could lead a clinician to misinterpret patient results
Falsely Elevated Results Observed with Plasma Specimens, could lead a clinician to misinterpret patient results
Potential for falsely elevated hCG results due to sample carryover in serum and urine patient samples, as well as quality control samples and adjustors.
Dimension Vista Magnesium Flex reagent cartridge lots exhibiting imprecision and intermittent Abnormal Assay flags. When this issue occurs, there is a potential for QC failures or erroneous…
(1)Software (SW) versions V1.25.1 and lower may result in TDef (Test Definition) Parameters for Open Channel Assays reverting to Default Value would shift test results by the magnitude of…
¿2 Microglobulin (B2M) Reagent Lot 211446 Bias with Quality Control and Patient Sample Results. Positive bias on patient samples of up to +20.9% at a B2M concentration of approximately 1.5…
Incorrect Default Hemolysis, Icterus or Lipemia (HIL) Index for Five Dimension Vista Assays
Incorrect Default Hemolysis, Icterus or Lipemia (HIL) Index for Five Dimension Vista Assays
Data provided in the Emergency Use Authorization (EUA) submission was insufficient to determine that the assays were adequately validated for the intended use.
Data provided in the Emergency Use Authorization (EUA) submission was insufficient to determine that the assays were adequately validated for the intended use.
The Aptio Automation Firmware for the ADVIA Centaur XP/XPT Interface Module may lead, in some cases, to an incorrect association of test results to sample ID. Incorrect association of test…
The firm has confirmed the potential for ADVIA Chemistry Urinary/Cerebrospinal Fluid Protein reagent carryover impacting Enzymatic Hemoglobin A1c (A1c_E and A1c_EM) results. Falsely…
The firm has confirmed the potential for Atellica CH Urinary/Cerebrospinal Fluid Protein reagent carryover impacting Enzymatic Hemoglobin A1c (A1c_E and A1c_EM) results. Falsely depressed…
Ancillary reagent packs used on Atellica IM 1300 and Atellica IM 1600 analyzers may overflow when pierced while on an analyzer located at an altitude of greater than 350m (1148ft) above sea…
Ancillary reagent packs used on Atellica IM 1300 and Atellica IM 1600 analyzers may overflow when pierced while on an analyzer located at an altitude of greater than 350m (1148ft) above sea…
Product does not meet the Instructions for Use (IFU) claimed Pack Calibration Interval of 10 days. IFU calibration updated interval from 10 days to 3 days for all current and future kit lot…
Product does not meet the Instructions for Use (IFU) claimed Pack Calibration Interval of 10 days. IFU calibration updated interval from 10 days to 3 days for all current and future kit lot…
Product does not meet the Instructions for Use (IFU) claimed Pack Calibration Interval of 10 days. IFU calibration updated interval from 10 days to 3 days for all current and future kit lot…
Valves used on the Atellica CH 930, Atellica IM 1300 and Atellica IM 1600 analyzers may develop a malfunction due to a manufacturing defect and may result in the valve wearing and leaking…
ADVIA Chemistry System and Software Versions (V1.4 SMN 11314625 and V1.4 China SMN 11316885) may not be aligned with customer expectations for the current default mixer failure detection…
Valves used on the Atellica CH 930, Atellica IM 1300 and Atellica IM 1600 analyzers may develop a malfunction due to a manufacturing defect and may result in the valve wearing and leaking…
Valves used on the Atellica CH 930, Atellica IM 1300 and Atellica IM 1600 analyzers may develop a malfunction due to a manufacturing defect and may result in the valve wearing and leaking…
This page summarizes recall records in which the official source identifies Siemens Healthcare Diagnostics, Inc. as the recalling firm or manufacturer. The number of records should not be treated as a safety rating or a measure of company size, because recall counts depend on product mix, reporting periods and how government records are organized.
Compare the dates, categories and classifications together rather than relying on the total alone. Open an individual record for the recalled product, reason, distribution details and current instructions available from the source.