Software anomalies were reported to affect the following system functionalities: startup and shutdown, system and network communication, motion control, image acquisition and output, dose…
Class IIFDA: TerminatedMedtronic Navigation, Inc.-Littleton
The pouches of certain lots of product may have wrinkles along the pouch seal that could result in a channel within the seal. Should there be a channel in the pouch seal it may result in a…
The BACT/ALERT VIRTUO Instrument Firmware version R2.0 allows relocation of the four (4) resident calibration standards so that Field System Engineers (FSE) can service the associated cells…
The system's IFU contains a list of generators that the have been deemed compatible. Recent reports indicate that one energy platform is not compatible with the system. Therefore, it has…
Class IIFDA: TerminatedTerumo Cardiovascular Systems Corporation
Software anomalies were reported to affect the following system functionalities: startup and shutdown, system and network communication, motion control, image acquisition and output, dose…
Class IIFDA: TerminatedMedtronic Navigation, Inc.-Littleton
Software anomalies were reported to affect the following system functionalities: startup and shutdown, system and network communication, motion control, image acquisition and output, dose…
Class IIFDA: TerminatedMedtronic Navigation, Inc.-Littleton
Stitching run on SkyPlate detector interrupted intermittently; Stitching with SkyPlate aborts after first image; System software CD contains wrong baseline
Class IIFDA: TerminatedPhilips Medical Systems Gmbh, DMC
GE Healthcare has recently become aware of a potential safety issue with the 3.0T 6 Channel Flex Coil used with the 3T MR750W Surgical Suite Scanners. Coil overheating can occur when the…
The probe power monitoring settings have been found to be not appropriate under specific operating conditions when connected to the devices listed below. This issue may result in…
Reprocessing instructions for the SGS/SGA Surgical Handpieces have been updated with instructions of manual cleaning and specific parameters for autoclave sterilization . The previous…
Reprocessing instructions for the SGS/SGA Surgical Handpieces have been updated with instructions of manual cleaning and specific parameters for autoclave sterilization. The previous…
The probe power monitoring settings have been found to be not appropriate under specific operating conditions when connected to the devices listed below. This issue may result in…
A small area on the bore surface, which is normally accessible to the patient, can get warm enough to cause a serious burn in the event the patient touches the bore and proper padding is…
When using respiratory gating with Steady Mode ON in Spin Echo sequence scanning, the SAR (Specific Absorption Rate) value displayed on the console may be lower than the actual SAR value.…
Class IIFDA: OngoingCanon Medical System, USA, Inc.
Analysis of the completed dataset from the COMPASS-XT long-term study showed that the CyPass Micro-Stent group experienced statistically significant endothelial cell loss (ECL) compared to…
The probe power monitoring settings have been found to be not appropriate under specific operating conditions when connected to the devices listed below. This issue may result in…
Software anomalies were reported to affect the following system functionalities: startup and shutdown, system and network communication, motion control, image acquisition and output, dose…
Class IIFDA: TerminatedMedtronic Navigation, Inc.-Littleton
Analysis of the completed dataset from the COMPASS-XT long-term study showed that the CyPass Micro-Stent group experienced statistically significant endothelial cell loss (ECL) compared to…
The pouches of certain lots of product may have wrinkles along the pouch seal that could result in a channel within the seal. Should there be a channel in the pouch seal it may result in a…