FDA recall record

Canon Medical System, USA Nuclear Magnetic Resonance Imaging Systems that are indicated for recall — FDA Z-0199-2019

When using respiratory gating with Steady Mode ON in Spin Echo sequence scanning, the SAR (Specific Absorption Rate) value displayed on the console may be…

Class IIclassification
Z-0199-2019official record ID
Oct 31, 2018report date

Recall record at a glance

Official recordZ-0199-2019
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusOngoing
Recalling firmCanon Medical System, USA, Inc.
Report dateOct 31, 2018
Recall initiation dateMay 30, 2018
Quantity252
LocationTustin, CA, United States

Official product description

Nuclear Magnetic Resonance Imaging Systems that are indicated for use as a diagnostic imaging modality that produces cross-sectional transaxial, coronal, sagittal, and oblique images that display anatomic structures of the head or body.

Why the product was recalled

When using respiratory gating with Steady Mode ON in Spin Echo sequence scanning, the SAR (Specific Absorption Rate) value displayed on the console may be lower than the actual SAR value. The actual applied SAR values may exceed the FIRST LEVEL CONTROLLED OPERATING MODE.

Distribution information

Domestic: AL, AR, AZ, CA, CO, CT, FL, GA, IA, ID, IL, IN, KS, LA, MA, MD, MI, MN, MO, MS, MT, NC, NH, NJ, NM, NV, NY, OH, OK, OR, PA, SC, SD, TN, TX, UT, VT, WA, WI, WV, and WY. International: Algeria, Australia, Bangladesh, Brazil, Canada, China, Colombia, Congo, Dominican Republic, Egypt, Guinea, Honduras, India, Indonesia, Iran, Japan, Kenya, Korea, Lebanon, Malaysia, Mexico, Morocco, Netherlands, Nigeria, Oman, Pakistan, Paraguay, Qatar, Saudi Arabia, Singapore, South Africa, Sudan, Taiwan, Thailand, Turkey, UAE, and Vietnam

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.