GE Medical Systems GE Healthcare, Discovery 3.0T MR750, 5148810-2, 5373011-2, 5499460-2 recall — FDA Z-0201-2019

GE Healthcare has recently become aware of a potential safety issue with the 3.0T 6 Channel Flex Coil used with the 3T MR750W Surgical Suite Scanners.…

Class IIclassification
Z-0201-2019official record ID
Oct 31, 2018report date

Recall record at a glance

Official recordZ-0201-2019
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusTerminated
Recalling firmGE Medical Systems, LLC
Report dateOct 31, 2018
Recall initiation dateDec 7, 2015
Quantity356 (98 US; 258 OUS)
LocationWaukesha, WI, United States

Official product description

GE Healthcare, Discovery 3.0T MR750, 5148810-2, 5373011-2, 5499460-2.

Why the product was recalled

GE Healthcare has recently become aware of a potential safety issue with the 3.0T 6 Channel Flex Coil used with the 3T MR750W Surgical Suite Scanners. Coil overheating can occur when the device is used in Mode 2 setup. This could lead to a serious patient thermal injury. There have been no injuries reported as a result of this issue. UPDATE 5/27/2016: Recall has been updated to include an additional product which includes the Discovery 3.0T MR750 systems running Application Software DV24 R01, DV25 R02, or DV25.1 M3.

Distribution information

Worldwide Distribution- US Distribution including the states of AZ, CA, CO, IL IN, MI, MN, NY, OH, OK, SD, TX, WI, VA, DE, FL, MD, ME, MO, NC, NH, OR, PA, TN, WI, DC, PR and the countries of CANADA, CHINA, CZECH REPUBLIC, ITALY, JAPAN, NETHERLANDS, NEW ZEALAND, PORTUGAL, ROMANIA, SOUTH AFRICA, SWITZERLAND, TURKEY, SWEDEN, AUSTRALIA, BRAZIL, CZECH REPUBLIC, EGYPT, FRANCE, GERMANY, ITALY, MEXICO, RUSSIA, SAUDI ARABIA, TAIWAN, UNITED KINGDOM, INDONESIA, ESTONIA, REPUBLIC OF KOREA, POLAND, KUWAIT, SPAIN.

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Telephone.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.

Related recall records

Other records from the same recalling company are shown first when available.