FDA recall record

GE Medical Systems SIGNA Architect system Product Usage: The Signa Architect system is a recall — FDA Z-0236-2019

A small area on the bore surface, which is normally accessible to the patient, can get warm enough to cause a serious burn in the event the patient…

Class IIclassification
Z-0236-2019official record ID
Oct 31, 2018report date

Recall record at a glance

Official recordZ-0236-2019
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusTerminated
Recalling firmGE Medical Systems, LLC
Report dateOct 31, 2018
Recall initiation dateOct 3, 2018
Quantity4 units
LocationWaukesha, WI, United States

Official product description

SIGNA Architect system Product Usage: The Signa Architect system is a whole body magnetic resonance scanner designed to support high resolution, high signal-to-noise ratio, and short scan times. It is indicated for use as a diagnostic imaging device to produce axial, sagittal, coronal, and oblique images, spectroscopic images, parametric maps, and/or spectra, dynamic images of the structures and/or functions of the entire body, including, but not limited to, head, neck, TMJ, spine, breast, heart, abdomen, pelvis, joints, prostate, blood vessels, and musculoskeletal regions of the body. Depending on the region of interest being imaged, contrast agents may be used. The images produced reflect the spatial distribution or molecular environment of nuclei exhibiting magnetic resonance. These images and/or spectra when interpreted by a trained physician yield information that may assist in diagnosis.

Why the product was recalled

A small area on the bore surface, which is normally accessible to the patient, can get warm enough to cause a serious burn in the event the patient touches the bore and proper padding is not used.

Distribution information

Worldwide - US Nationwide Distribution in the state of Utah. International distribution to France, Republic of Korea, and Sweden.

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.