Diagnostica Stago STA Staclot Protein S recall — FDA Z-3047-2026
A positive drift in QC values within the reagent's stated 4-hour onboard stability period was identified
Overview of recall records associated with Diagnostica Stago, Inc., including dates, categories and classifications where available.
Classification mix: Class II: 19.
A positive drift in QC values within the reagent's stated 4-hour onboard stability period was identified
After receiving customer complaints, investigations confirmed the presence of a positive bias in D-Dimer measurements throughout the entire analytical range.
The potential of out-of-range results and an underestimation of the free protein S level in normal patient plasmas. The defect only affects normal-range values.
Potential for defect for Asserachrom HPIA lot 271288, which can affect every well in the impacted strip and produce negative results.
Potential for contamination of STA-LIQUID ANTI-Xa by Stago Fibrinogen reagents in UFH and LMWH applications.
Potential for contamination of STA-LIQUID ANTI-Xa by Stago Fibrinogen reagents in UFH and LMWH applications.
Potential for contamination of STA-LIQUID ANTI-Xa by Stago Fibrinogen reagents in UFH and LMWH applications.
Potential for contamination of STA-LIQUID ANTI-Xa by Stago Fibrinogen reagents in UFH and LMWH applications.
Potential for shorter clotting times in clotting tests which have a dilution, leading to possible overestimate of around 1%
An internal investigation identified a bug in this firmware version, resulting in the following:Intermittent shortened coagulation times and an increased frequency of occurrence of a…
An internal investigation identified a bug in this firmware version, resulting in the following:Intermittent shortened coagulation times and an increased frequency of occurrence of a…
An internal investigation identified a bug in this firmware version, resulting in the following:Intermittent shortened coagulation times and an increased frequency of occurrence of a…
A risk of sample-to-sample cross-contamination of Emicizumab for factor VIII assays on Stago instruments, clotting times will be shortened that may cause factor VIII levels reported to be…
A positive bias was confirmed to exists on the STA - Unicalibrator levels with STA - Deficient IX and STA - Deficient VIII when compared to the International Standard (IS) for factor VIII…
A positive bias was confirmed to exists on the STA - Unicalibrator levels with STA - Deficient IX and STA - Deficient VIII when compared to the International Standard (IS) for factor VIII…
There have been reports of shortened (T1-T2) results, where some instances of result recovery outside of the range of less than or equal to 15 seconds.
There have been reports of shortened (T1-T2) results, where some instances of result recovery outside of the range of less than or equal to 15 seconds.
QC values outside of the assigned ranges (prolonged Prothrombin time / decreased PT %)
QC values outside of the assigned ranges (prolonged Prothrombin time / decreased PT %)
This page summarizes recall records in which the official source identifies Diagnostica Stago, Inc. as the recalling firm or manufacturer. The number of records should not be treated as a safety rating or a measure of company size, because recall counts depend on product mix, reporting periods and how government records are organized.
Compare the dates, categories and classifications together rather than relying on the total alone. Open an individual record for the recalled product, reason, distribution details and current instructions available from the source.