FDA recall record

Diagnostica Stago STA Compact Automated Multi-Parametric Analyzer, Reference Numbers 58602 recall — FDA Z-1576-2023

An internal investigation identified a bug in this firmware version, resulting in the following:Intermittent shortened coagulation times and an increased…

Class IIclassification
Z-1576-2023official record ID
May 17, 2023report date

Recall record at a glance

Official recordZ-1576-2023
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusOngoing
Recalling firmDiagnostica Stago, Inc.
Report dateMay 17, 2023
Recall initiation dateMar 22, 2023
Quantity2 analyers
LocationParsippany, NJ, United States

Official product description

STA Compact Automated Multi-Parametric Analyzer, Reference Numbers 58602 and 58612

Why the product was recalled

An internal investigation identified a bug in this firmware version, resulting in the following:Intermittent shortened coagulation times and an increased frequency of occurrence of a technical error reported by the analyzer (error 13).

Distribution information

US Nationwide distribution in the states of IL, MA, MN, NE, TX, and WI.

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.

Related recall records

Other records from the same recalling company are shown first when available.

Medical DeviceMar 25, 2026

STA Liatest D-Di recall — FDA Z-1585-2026

After receiving customer complaints, investigations confirmed the presence of a positive bias in D-Dimer measurements throughout the entire analytical range.

Class IIFDA: OngoingDiagnostica Stago, Inc.