FDA recall record

Diagnostica Stago Staclot LA 20 (REF 00594) The Staclot LA and Staclot LA 20 test kits are recall — FDA Z-0611-2019

There have been reports of shortened (T1-T2) results, where some instances of result recovery outside of the range of less than or equal to 15 seconds.

Class IIclassification
Z-0611-2019official record ID
Dec 26, 2018report date

Recall record at a glance

Official recordZ-0611-2019
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusTerminated
Recalling firmDiagnostica Stago, Inc.
Report dateDec 26, 2018
Recall initiation dateNov 14, 2018
Quantity22342 total
LocationParsippany, NJ, United States

Official product description

Staclot LA 20 (REF 00594) The Staclot LA and Staclot LA 20 test kits are reagent systems designed for the qualitative detection of lupus anticoagulants (LA) in plasma by the use of hexagonal HII phase phospholipid molecules (1, 2).

Why the product was recalled

There have been reports of shortened (T1-T2) results, where some instances of result recovery outside of the range of less than or equal to 15 seconds.

Distribution information

US Nationwide Distribution

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.

Related recall records

Other records from the same recalling company are shown first when available.

Medical DeviceMar 25, 2026

STA Liatest D-Di recall — FDA Z-1585-2026

After receiving customer complaints, investigations confirmed the presence of a positive bias in D-Dimer measurements throughout the entire analytical range.

Class IIFDA: OngoingDiagnostica Stago, Inc.