Diagnostica Stago Asserachrom HPIA recall — FDA Z-1084-2026

Potential for defect for Asserachrom HPIA lot 271288, which can affect every well in the impacted strip and produce negative results.

Class IIclassification
Z-1084-2026official record ID
Jan 21, 2026report date

Recall record at a glance

Official recordZ-1084-2026
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusOngoing
Recalling firmDiagnostica Stago, Inc.
Report dateJan 21, 2026
Recall initiation dateDec 15, 2025
Quantity438 units
LocationParsippany, NJ, United States

Official product description

Asserachrom HPIA. Model/Catalog Number: 00615. The Asserachrom HPIA kit is intended for the qualitative detection in plasma or serum of the anti-heparin-platelet factor 4 (PF4) antibodies generated during the type II heparin-induced thrombocytopenia (HIT type II) by ELISA.

Why the product was recalled

Potential for defect for Asserachrom HPIA lot 271288, which can affect every well in the impacted strip and produce negative results.

Distribution information

Worldwide - US Nationwide distribution in the states of AL, CA, FL, GA, HI, IA, MA, MI, NJ, NV, NY, OH, TN, WI and the countries of Belgium, Canada, Denmark, Algeria, Spain, Finland, France, Great Britain, Italy, Lebanon, Saudi Arabia, Uganda.

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: E-Mail.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.

Related recall records

Other records from the same recalling company are shown first when available.

Medical DeviceMar 25, 2026

STA Liatest D-Di recall — FDA Z-1585-2026

After receiving customer complaints, investigations confirmed the presence of a positive bias in D-Dimer measurements throughout the entire analytical range.

Class IIFDA: OngoingDiagnostica Stago, Inc.