Medical DeviceNov 7, 2018

Siemens Medical Solutions USA Ysio recall — FDA Z-0299-2019

In rare cases of high clinical workload, the steel cables inside the lifting column of the overhead tube, which are designed to take the load, may rapture without triggering the safety…

Class IIFDA: TerminatedSiemens Medical Solutions USA, Inc.
Medical DeviceNov 7, 2018

Arrow AGB + Multi-Lumen CVC Kit, Cat recall — FDA Z-0282-2019

The product was shipped after its expiry date due to a system error. The product lidstock identifies the correct expiration date however the accompanying purchase order shipping…

Class IIFDA: TerminatedArrow International Inc.