FDA recall record

Roche Diagnostics URISYS 1100 Urine Analyzer Part Number: 03617556001 Product Usage: The recall — FDA Z-0301-2019

Limits of Detection (LoD) for protein, nitrite, leukocytes, and erythrocytes on the Urisys 1100 urine analyzer with Chemstrip 5 OB, Chemstrip 7, Chemstrip…

Class IIclassification
Z-0301-2019official record ID
Nov 7, 2018report date

Recall record at a glance

Official recordZ-0301-2019
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusTerminated
Recalling firmRoche Diagnostics Corporation
Report dateNov 7, 2018
Recall initiation dateSep 24, 2018
Quantity14,398 units
LocationIndianapolis, IN, United States

Official product description

Roche Diagnostics URISYS 1100 Urine Analyzer Part Number: 03617556001 Product Usage: The URISYS 1100 Urine Analyzer is a reflectance photometer designed to automatically read and evaluate the results of Chemstrip 5 OB, Chemstrip 7, Chemstrip 10 MD, and Chemstrip 10 UA urine test strips for various urine analytes.: specific gravity, pH, leukocytes, nitrite, protein, glucose, ketones, urobilinogen, bilirubin and blood

Why the product was recalled

Limits of Detection (LoD) for protein, nitrite, leukocytes, and erythrocytes on the Urisys 1100 urine analyzer with Chemstrip 5 OB, Chemstrip 7, Chemstrip 10 MD, and Chemstrip 10 UA test strips were higher than what is listed in the test strip method sheets may lead to false negative results for the affected parameters

Distribution information

US Nationwide Distribution

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.