FDA recall record

Siemens Medical Solutions USA Ysio MAX (model number 10762470) with Tube stand 3D V recall — FDA Z-0300-2019

In rare cases of high clinical workload, the steel cables inside the lifting column of the overhead tube, which are designed to take the load, may rapture…

Class IIclassification
Z-0300-2019official record ID
Nov 7, 2018report date

Recall record at a glance

Official recordZ-0300-2019
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusTerminated
Recalling firmSiemens Medical Solutions USA, Inc.
Report dateNov 7, 2018
Recall initiation dateAug 6, 2018
Quantity31
LocationMalvern, PA, United States

Official product description

Ysio MAX (model number 10762470) with Tube stand 3D V (model number 7042018 or 7042034), radiographic systems used in radiographic and tomographic exposures of the whole body.

Why the product was recalled

In rare cases of high clinical workload, the steel cables inside the lifting column of the overhead tube, which are designed to take the load, may rapture without triggering the safety lock. This could lead to overloading the safety rope, which is not designed for continuous load and ongoing movement under load. This may lead to mechanical fatigue and cause the arm to drop down unexpectedly during movement in vertical direction and during patient positioning.

Distribution information

The products were distributed to the following US states: AL, AR, AZ, CA, CO, DC, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MI, MN, MO, MS, NC, ND, NE, NM, NV, NY, OH, OK, PA, PR, SC, SD, TX, UT, VA, WI, and WY.

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.

Related recall records

Other records from the same recalling company are shown first when available.