FDA recall record

Biomerieux The VITEK 2 Gram-positive Susceptibility Card is intended for use with the VITEK recall — FDA Z-1742-2020

There is a potential for the occurrence of false positive OXSF results for some Staphylococcus aureus strains in conjunction with the VITEK 2 AST-P652…

Class IIclassification
Z-1742-2020official record ID
Apr 29, 2020report date

Recall record at a glance

Official recordZ-1742-2020
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusOngoing
Recalling firmBiomerieux Inc.
Report dateApr 29, 2020
Recall initiation dateMar 11, 2020
Quantity1698 cartons (15,940 cards)
LocationHazelwood, MO, United States

Official product description

The VITEK 2 Gram-positive Susceptibility Card is intended for use with the VITEK 2 Systems in clinical laboratories as an in-vitro test to determine susceptibility of Staphylococcus spp., Enterococcus spp., and S. agalactiae to antimicrobial agents when used as instructed.

Why the product was recalled

There is a potential for the occurrence of false positive OXSF results for some Staphylococcus aureus strains in conjunction with the VITEK 2 AST-P652 test card (ref. 421857, lot 8021232403). .

Distribution information

No distribution within the United States. Foreign distribution locations: Netherlands, Luxembourg, Belgium

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.

Related recall records

Other records from the same recalling company are shown first when available.

Medical DeviceMay 1, 2024

Biomerieux VITEK 2 - AST Test kits recall — FDA Z-1464-2024

Ceftriaxone (cro02n) concentration errors associated with specific recently manufactured test kits which would have the potential for for false susceptible results. The risk is present for…

Class IFDA: OngoingBiomerieux Inc.