bioMerieux VITEK 2 Gram-negative AST Test Kits (Cards) containing Colistin formulation cs02n recall — FDA Z-2671-2025

Potential for false resistant results for Colistin (cs02n) with multi-drug-resistant Pseudomonas aeruginosa and Acinetobacter baumannii complex used in…

Class IIclassification
Z-2671-2025official record ID
Oct 8, 2025report date

Recall record at a glance

Official recordZ-2671-2025
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusOngoing
Recalling firmbioMerieux, Inc.
Report dateOct 8, 2025
Recall initiation dateJun 26, 2025
Quantity235269 cards
LocationDurham, NC, United States

Official product description

VITEK 2 Gram-negative AST Test Kits (Cards) containing Colistin formulation cs02n. VITEK 2 is an automated system consisting of instruments, software and reagent cards designed for the identification and antimicrobial susceptibility testing of bacteria and yeast. Export only

Why the product was recalled

Potential for false resistant results for Colistin (cs02n) with multi-drug-resistant Pseudomonas aeruginosa and Acinetobacter baumannii complex used in specific cards

Distribution information

International distribution in the countries of Albania, Armenia, Australia, Austria, Belgium, Bosnia, Bulgaria, China, Croatia, Cyprus, Czech Republic/Slovakia, France, Germany, Greece, Iberia, India, Israel, Italy, Japan, Lithuania, Macedonia, Malaysia/Brunei, Moldova, Netherlands, Pakistan, Poland, Portugal, Romania, Russia, Serbia, South Korea, Spain, Switzerland, Taiwan, Thailand, and Turkey.

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.