Biomerieux VITEK 2 - AST Test kits recall — FDA Z-1464-2024

Ceftriaxone (cro02n) concentration errors associated with specific recently manufactured test kits which would have the potential for for false…

Class Iclassification
Z-1464-2024official record ID
May 1, 2024report date

Recall record at a glance

Official recordZ-1464-2024
CategoryMedical Device Recalls
ClassificationClass I
FDA record statusOngoing
Recalling firmBiomerieux Inc.
Report dateMay 1, 2024
Recall initiation dateMar 14, 2024
Quantity49215 units
LocationHazelwood, MO, United States

Official product description

VITEK 2 - AST Test kits (in vitro diagnostic test cards for antimicrobial susceptibility): a) AST-N334, REF 418984; b) AST-GN95, REF 421982; c) AST-GN99, REF 423102; d) AST-N390, REF 423340; e) AST-N391, REF 423341; f) AST-N392, REF 423342; g) AST-N395, REF 423491; h) AST-N401, REF 423643; i) AST-N402, REF 423644; j) AST-N404, REF 423664; k) AST-N802, REF 423706; l) AST-N405, REF 423864; m) AST-N417, REF 423880; n) AST-N408, REF 423924; o) AST-N409, REF 423925; p) AST-N420, REF 424039; q) AST-N423, REF 424042; r) AST-N422, REF 424056; s) AST-N436, REF 424440; t) AST-N809, REF 424703; u) AST-N806, REF 424709; v) AST-N807, REF 424710; w) AST-N808, REF 424711; x) AST-N810, REF 424712; y) AST-N812, REF 424721

Why the product was recalled

Ceftriaxone (cro02n) concentration errors associated with specific recently manufactured test kits which would have the potential for for false susceptible results. The risk is present for isolates that have a result of MIC=0.5, 1, 2.

Distribution information

Worldwide

How to read the classification

FDA Class I indicates a situation in which there is a reasonable probability that use of or exposure to the product will cause serious adverse health consequences or death. The classification shown here comes directly from the FDA enforcement record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: E-Mail.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.

Related recall records

Other records from the same recalling company are shown first when available.