Recall record at a glance
| Official record | Z-1926-2013 |
|---|---|
| Category | Medical Device Recalls |
| Classification | Class II |
| FDA record status | Terminated |
| Recalling firm | SpineFrontier, Inc. |
| Report date | Aug 14, 2013 |
| Recall initiation date | Mar 14, 2013 |
| Quantity | 8 |
| Location | Beverly, MA, United States |
Official product description
Rev E PedFuse Return Screw Inserter, Part Number: 11-80030 Screw inserters are instruments which capture, hold and drive pedicle screw assemblies. There is a capturing feature that engages and holds the screw in place. There is also a feature used to lock the screw to the inserter before insertion of the screw. They can also be used to remove screws or back screws out. Screw inserters are used during the implantation of spinal pedicle screws intended to provide immobilization and stabilization of spinal segments in skeletally mature patients. Used in conjunction with Pedicle Screw Spinal system marketed as KRD1 PedFuse Pedicle Screw System .
Why the product was recalled
Inserters have a set screw that fixes a collar to the inserter. The screw set can loosen and fall out of the inserter which allows the collar to slide off the proximal end of the inserter. When the collar is not in place the locking sleeve can also slide off the inserter.
Distribution information
Distribution only in Texas.
How to read the classification
FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.
Additional record details
Recall type: Voluntary: Firm initiated. Initial notification: E-Mail.
What to do next
If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.
About this record
Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.
FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.