Recall record at a glance
| Official record | Z-0777-2014 |
|---|---|
| Category | Medical Device Recalls |
| Classification | Class II |
| FDA record status | Terminated |
| Recalling firm | SpineFrontier, Inc. |
| Report date | Jan 22, 2014 |
| Recall initiation date | Sep 24, 2013 |
| Quantity | 23 devices in total |
| Location | Beverly, MA, United States |
Official product description
Arena-L 35x26 Trial Head I 10 Lordotic; Part: 11-81004-08, 11-81004-10, 11-81004-12, 11-81004-14, 11-81004-16 The trial heads are smooth plates that mimic the footprint of the implant, manufactured from 17-4 Stainless Steel into 35X26mm (11-81004-XX) and 38X28mm (11-81012-XX) varieties, each with multiple height configurations. The Trial Heads are intended to attach tot he Trial Handle (11-81002) and to be inserted into the intervertebral disc space. The Trial Handle consists of a knurled handle and cylindrical shaft with an M4 threads interface at the distal end; 17-4 Stainless steel is used as the material for the handles. The Trial Heads distract the disc space to the surgeon's desired height to gauge which size implant should be utilized to complete the Anterior Lumbar Interbody Fusion Procedure.
Why the product was recalled
Handles and heads may not be compatible.
Distribution information
US Distribution: VA only.
How to read the classification
FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.
Additional record details
Recall type: Voluntary: Firm initiated. Initial notification: E-Mail.
What to do next
If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.
About this record
Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.
FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.