FDA recall record

SpineFrontier S-Lift Lock Shaft Inserter, an instrument used with S-LIFT Intervertebral Body recall — FDA Z-1565-2014

It has been verified through field use that there are two possible breakages that may occur when using SI50001 Revision D Lock Shaft Inserter, possibly…

Class IIclassification
Z-1565-2014official record ID
May 14, 2014report date

Recall record at a glance

Official recordZ-1565-2014
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusTerminated
Recalling firmSpineFrontier, Inc.
Report dateMay 14, 2014
Recall initiation dateDec 23, 2011
Quantity30 devices
LocationBeverly, MA, United States

Official product description

SpineFrontier S-Lift Lock Shaft Inserter, an instrument used with S-LIFT Intervertebral Body Fusion Device in spinal fusion surgery. Product Usage: The S-Lift Lock Shaft SI50001 mates with SI50002, Implant Inserter, to aid in securing the implant to the whole implant inserter assembly, SI50000.

Why the product was recalled

It has been verified through field use that there are two possible breakages that may occur when using SI50001 Revision D Lock Shaft Inserter, possibly resulting in instrument fragments entering the sterile field: Impact Cap: The impact cap on the Lock Shaft Inserter can shear off under high-impact force during insertion and removal. Pin: The pin used to guide the Lock Shaft Inserter into the ma

Distribution information

US Nationwide - OR, FL, MA, TX, CO, OK, IL, and MO

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.

Related recall records

Other records from the same recalling company are shown first when available.