FDA recall record

Topcon Medical Laser Systems PASCAL and PASCAL Streamline (532 nm.577 nm) recall — FDA Z-1774-2013

When using the "Favorites" function for saved settings, there is potential for a mismatch between displayed power and the actual power output. This may…

Class IIclassification
Z-1774-2013official record ID
Jul 31, 2013report date

Recall record at a glance

Official recordZ-1774-2013
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusTerminated
Recalling firmTopcon Medical Laser Systems, Inc.
Report dateJul 31, 2013
Recall initiation dateMar 27, 2012
Quantity7 US, 33 foreign
LocationSanta Clara, CA, United States

Official product description

PASCAL and PASCAL Streamline (532 nm.577 nm); Treats the patient's eye.

Why the product was recalled

When using the "Favorites" function for saved settings, there is potential for a mismatch between displayed power and the actual power output. This may result in over or under treatment.

Distribution information

Worldwide Distribution-USA and the countries of China, Hong Kong, Brazil, France, Japan, Netherlands, Pakistan, Malaysia, United Kingdom, South Korea, Thailand, Dubai, and South Africa.

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.

Related recall records

Other records from the same recalling company are shown first when available.