FDA recall record

Topcon Medical Laser Systems Opthalmic: PASCAL Synthesis is an ophthalmic scanning laser recall — FDA Z-2405-2016

Topcon discovered an anomaly with titration, only when EpM is turned on during the titration process. This event may occur if the user turns on EpM, turns…

Class IIclassification
Z-2405-2016official record ID
Aug 17, 2016report date

Recall record at a glance

Official recordZ-2405-2016
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusTerminated
Recalling firmTopcon Medical Laser Systems, Inc.
Report dateAug 17, 2016
Recall initiation dateJul 15, 2016
Quantity205 units (21 in US, 184 OUS)
LocationSanta Clara, CA, United States

Official product description

Opthalmic: PASCAL Synthesis is an ophthalmic scanning laser system. The system can perform single shot photocoagulation as is performed conventionally.

Why the product was recalled

Topcon discovered an anomaly with titration, only when EpM is turned on during the titration process. This event may occur if the user turns on EpM, turns off the Landmark (LM) function (with any EpM supported pattern) and then enters titration mode.

Distribution information

Worldwide Distribution: US (nationwide) and countries of: Canada, Mexico, Japan, Netherlands, Algeria, Australia, Brazil, Chile, China, France, Guatemala, Hong Kong, Italy, New Zealand, Pakistan, Philippines, Singapore, Taiwan, and United Kingdom.

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.