Topcon Medical Laser Systems Opthalmic: PASCAL Synthesis is an ophthalmic scanning laser recall — FDA Z-2405-2016
Topcon discovered an anomaly with titration, only when EpM is turned on during the titration process. This event may occur if the user turns on EpM, turns off the Landmark (LM) function…