FDA recall record

Sentinel CH SPA MULTIGENT Creatinine (Enzymatic), Catalog Numbers 8L24-31 and 8L24-41 recall — FDA Z-2867-2017

Reagent 1 component may result in Quality Control values shifting out of acceptable range, which is not resolved by recalibration.

Class IIclassification
Z-2867-2017official record ID
Aug 9, 2017report date

Recall record at a glance

Official recordZ-2867-2017
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusTerminated
Recalling firmSentinel CH SPA
Report dateAug 9, 2017
Recall initiation dateMay 22, 2017
Quantity2356 kits Worldwide; US 206 kits
LocationMilan, N/A, Italy

Official product description

MULTIGENT Creatinine (Enzymatic), Catalog Numbers 8L24-31 and 8L24-41; IVD. Intended for the quantitative determination of creatinine in human serum, plasma, and urine using the ARCHITECT c Systems The product is a medical device (chemistry reagent) and is intended for the quantitative determination of creatinine in human serum, plasma, and urine using the ARCHITECT c Systems. The kit is provided as a ready to use liquid material consisting in two components: Reagent 1 and Reagent 2. According to the catalogue number, the kit is composed of three vials of 48 ml for Reagent 1 and three vials of 18 ml for Reagent 2 in the configuration 8L24-31. The configuration 8L24-41 consists in ten vials of 90 ml for Reagent 1 and ten vials of 34 ml for Reagent 2

Why the product was recalled

Reagent 1 component may result in Quality Control values shifting out of acceptable range, which is not resolved by recalibration.

Distribution information

Nationwide Distribution

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.