FDA recall record

Sentinel CH SpA Alinity c Lithium Reagent Kit- intended for the quantitation of lithium in recall — FDA Z-2552-2019

Added an additional SmartWash parameter for the Lithium and Alinity c Lithium Reagent Kit, catalogue number 8L25-30 and 08P5320 to prevent the potential…

Class IIclassification
Z-2552-2019official record ID
Oct 2, 2019report date

Recall record at a glance

Official recordZ-2552-2019
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusTerminated
Recalling firmSentinel CH SpA
Report dateOct 2, 2019
Recall initiation dateJun 28, 2019
Quantity1217 kits
LocationMilan, Italy

Official product description

Alinity c Lithium Reagent Kit- intended for the quantitation of lithium in serum or plasma using respectively the ARCHITECT c Systems and the Alinity c analyzer List Number: 08P5320

Why the product was recalled

Added an additional SmartWash parameter for the Lithium and Alinity c Lithium Reagent Kit, catalogue number 8L25-30 and 08P5320 to prevent the potential for carryover from the ARCHITECT Lactate Dehydrogenase (LDH) reagent (LN 2P56) may result infalsely elevated lithium patient results

Distribution information

Nationwide Foreign: ARGENTINA AUSTRALIA AUSTRIA AZERBAIJAN BANGLADESH BELARUS BELGIUM BOSNIA AND HERZEGOVI BRAZIL BRUNEI CANADA CHILE COLOMBIA COSTA RICA CROATIA CZECH REPUBLIC DENMARK FAROE ISLANDS FINLAND FRANCE GERMANY GREECE GREENLAND HONG KONG IRELAND ISRAEL ITALY JORDAN KUWAIT LEBANON LUXEMBOURG MALAYSIA MALDIVES MEXICO MOROCCO NETHERLANDS NEW ZEALAND NIGERIA NORWAY OMAN PAKISTAN PANAMA PARAGUAY PHILIPPINES POLAND PORTUGAL PUERTO RICO ROMANIA RWANDA SAUDI ARABIA SLOVAKIA SOUTH AFRICA SPAIN ST. VINCENT SWEDEN SWITZERLAND TAIWAN THAILAND TURKEY UAE UNITED KINGDOM URUGUAY

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: E-Mail.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.

Related recall records

Other records from the same recalling company are shown first when available.