FDA recall record

Sentinel CH SpA Alinity c CRP Vario Cardiac High Sensitivity Calibrator Kit (ref 07P5605) recall — FDA Z-0033-2022

Calibrator vials were incorrectly labeled. Use of incorrectly labeled vials will cause the instrument to generate a warning that calibrators are…

Class IIIclassification
Z-0033-2022official record ID
Oct 13, 2021report date

Recall record at a glance

Official recordZ-0033-2022
CategoryMedical Device Recalls
ClassificationClass III
FDA record statusTerminated
Recalling firmSentinel CH SpA
Report dateOct 13, 2021
Recall initiation dateJul 29, 2021
Quantity32
LocationMilan, Italy

Official product description

Alinity c CRP Vario Cardiac High Sensitivity Calibrator Kit (ref 07P5605) - Product Usage: use in the calibration of the Alinity c CRP Vario assay, Cardiac CRP high sensitivity method [Cardiac CRP], on the Alinity c analyzer.

Why the product was recalled

Calibrator vials were incorrectly labeled. Use of incorrectly labeled vials will cause the instrument to generate a warning that calibrators are incorrectly loaded and calibration will not be completed successfully.

Distribution information

US Nationwide distribution in the states of LA, MD, MN, ND, NY, WV, and Puerto Rico.

How to read the classification

FDA Class III generally covers situations in which use of or exposure to the product is not likely to cause adverse health consequences. The classification displayed here is taken from the FDA enforcement record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: not stated.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.

Related recall records

Other records from the same recalling company are shown first when available.