Recall record at a glance
| Official record | Z-1737-2013 |
|---|---|
| Category | Medical Device Recalls |
| Classification | Class II |
| FDA record status | Terminated |
| Recalling firm | SpineFrontier, Inc. |
| Report date | Jul 24, 2013 |
| Recall initiation date | Dec 30, 2011 |
| Quantity | 2 rods |
| Location | Beverly, MA, United States |
Official product description
KRD1 PedFuse Rods, Model: IM80100-13 Rev B Product Usage: Pedicle rods are cylindrical rods which act as a linking mechanism between pedicle screws. The rod sits with the pedicle screw tulip and they are locked into place with a set screw or locking cap. The rod may be placed along several pedicle screws. Component rods are used during implantation of pedicle screw system devices intended to provide immobilization and stabilization of spinal segments of skeletally mature patients.
Why the product was recalled
One of the rods in each of two test constructs did not meet fatigue requirement as outlined in ASTM F1717-04 dynamic compression testing 5 million cycles.
Distribution information
USA Nationwide Distribution including the state of TX
How to read the classification
FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.
Additional record details
Recall type: Voluntary: Firm initiated. Initial notification: Telephone.
What to do next
If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.
About this record
Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.
FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.