FDA recall record

Greatbatch Medical Coaxial Micro-Introducer Kit recall — FDA Z-1614-2014

Greatbatch identified that the documentation does not support the five-year shelf life of the Bard Access Division Coaxial Micro-Introducer Set.

Class IIclassification
Z-1614-2014official record ID
May 21, 2014report date

Recall record at a glance

Official recordZ-1614-2014
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusTerminated
Recalling firmGreatbatch Medical
Report dateMay 21, 2014
Recall initiation dateApr 28, 2014
Quantity6,642 sets (66,420 devices)
LocationMinneapolis, MN, United States

Official product description

Coaxial Micro-Introducer Kit. Greatbatch Part Number/Bard Access Systems Order Number/Part Description 10636-001 8004011 Kit Coax 10 Pack 4FR Bard 10636-002 8005011 Kit Coax 10 Pack 5FR Bard 10636-007 8004012 Kit Coax 10 Pack 4FR Bard 10636-008 8005012 Kit Coax 10 Pack 5FR Bard 10636-009 8004022 Kit Coax 10 Pack 5FR Bard 10636-010 8005022 Kit Coax 10 Pack 5FR Bard The Coaxial Micro-Introducer set contains a 21 gauge disposable introducer needle, 0.018 inch floppy tip guidewire, and a radiopaque Coaxial Micro Introducer consisting of a sheath and dilator. The Coaxial Micro-Introducer Set is used to introduce up to a 0.038 inch guidewire or catheter into the vascular system following a small 21 gauge needle stick.

Why the product was recalled

Greatbatch identified that the documentation does not support the five-year shelf life of the Bard Access Division Coaxial Micro-Introducer Set.

Distribution information

US Nationwide Distribution in the state of UT.

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.

Related recall records

Other records from the same recalling company are shown first when available.