Recall record at a glance
| Official record | Z-1190-2017 |
|---|---|
| Category | Medical Device Recalls |
| Classification | Class II |
| FDA record status | Terminated |
| Recalling firm | Greatbatch Medical |
| Report date | Feb 22, 2017 |
| Recall initiation date | Oct 19, 2016 |
| Quantity | 1,441 |
| Location | Plymouth, MN, United States |
Official product description
Offset Reamer Handle, T5766 Product Usage: The reamer handles are re-usable surgical instruments used in arthroplasty surgery of the hip joints, where either total joint replacement or resurfacing surgery is performed using a reamer cutting instrument. The reamer handles are intended for use with a surgical driver (active device). The reamer handle is connected at the distal end to a reamer. The T5766 reamer handle is an offset design with a drive shaft mechanism that spins on roller bearings and is encapsulated in an inner clam shell housing that is constrained within a Radel tube and outer housing.
Why the product was recalled
On June 14, 2016, Greatbatch discovered a significant increase in the number of field complaints for the T5766 Offset Reamer Handle in which the U-joints of the drive chain component seizes during orthopedic procedures rendering the device non-functional. The 2016 failure rate for this failure exceeds the occurrence rate as defined in the DFMEA.
Distribution information
Worldwide Distribution - US Nationwide in the states of IN and the countries of Canada, Austria, Germany, Japan.
How to read the classification
FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.
Additional record details
Recall type: Voluntary: Firm initiated. Initial notification: Letter.
What to do next
If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.
About this record
Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.
FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.