Greatbatch Medical Greatbatch, Offset Cup Impactor with the following Parts and Catalogue # recall — FDA Z-0367-2017

On December 17, 2015, Greatbatch discovered that the Standard Offset Cup Impactor with a POM-C handle grip failed sterility testing when housed in a…

Class Iclassification
Z-0367-2017official record ID
Dec 7, 2016report date

Recall record at a glance

Official recordZ-0367-2017
CategoryMedical Device Recalls
ClassificationClass I
FDA record statusTerminated
Recalling firmGreatbatch Medical
Report dateDec 7, 2016
Recall initiation dateAug 2, 2016
Quantity4478 (2906 US, 1572 OUS)
LocationPlymouth, MN, United States

Official product description

Greatbatch, Offset Cup Impactor with the following Parts and Catalogue #: Tl0600 ~ MMP3100CHA04 T10491 MIMP3100CHA04 T12230 M2M1P3100CHA04 T15752 ~MMP3100CHA04 T15949 ~MMP3100CHA04 T17238 MMP3100CHA04 T9954 MMP3100CHA04 T9955 MMP3100CHA04 T8468 MMP3100CHA04 T11506 MMP3100CHA13 T10281 MMP3100CHA04 T8044 ~MMP3100CHA04 T17203 87-4 20-00980 T8177 ~ MMP3100CHA04 T16934 MMP3100CHA62 Tll340 SHP9109.00010 T8043 ~ MMP3100CHA04 T17321 MMPP3100CHA04 T15311 GG920010029 T1 7650 255000110 T8143 803-t 1-103 Tl4384 EK-00200690 T1666t XEB227242 T9316 MIMP3100CHA04 Tl3642 500325 T10753 ~MMP3100CH04 T9348 ~ MMP3100CHA04 T13480 MMP3100CHA042 T10243 ~MIMP3001CHA08 Tll209 ~ MMP300CHA12 Tl0861 MMP3100CHA04 T16427 MMP3100CHA04 T9894 :-. MMP3100CHA04 T12861 MMP3100CHA\041 T8088 2002CUPI T1661 1 T16611 Standard Offset Cup Impactors are handheld, reusable surgical instruments used to position and seat acetabular cups into the reamed acetabulum during hip joint replacement surgery. They are equipped with a suitable connection system to the acetabular cup (e.g. a thread) and can be supplied with a fixed or removable shock absorber. A hammer is used to deliver the necessary impact to an anvil (or impactor disc) at the rear end of the instrument.

Why the product was recalled

On December 17, 2015, Greatbatch discovered that the Standard Offset Cup Impactor with a POM-C handle grip failed sterility testing when housed in a dedicated instrument case using the sterilization parameters identified in the Standard Offset Cup Impactors Instructions for Use (IFU).

Distribution information

Worldwide Distribution - US including CA, IN, KS,NJ, NC, TN, TX, and UT and Internationally to ITALY, JAPAN, FRANCE, SPAIN, GERMANY, NETHERLANDS, GREECE, LUXEMBOURG, SOUTH KOREA, IRELAND, UNITED KINGDOM, AUSTRALIA, AUSTRIA, BELGIUM, SWITZERLAND, NORWAY, SWEDEN, and TAIWAN.

How to read the classification

FDA Class I indicates a situation in which there is a reasonable probability that use of or exposure to the product will cause serious adverse health consequences or death. The classification shown here comes directly from the FDA enforcement record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.