Radiometer ABL800 analyzer with FLEXQ module recall — FDA Z-0231-2018

Due to misinterpretation of the barcode by the scanner, when the registration receipt barcode is scanned by the analyzer , a result from a different…

Class IIclassification
Z-0231-2018official record ID
Dec 20, 2017report date

Recall record at a glance

Official recordZ-0231-2018
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusTerminated
Recalling firmRadiometer America Inc.
Report dateDec 20, 2017
Recall initiation dateOct 24, 2017
Quantity1,749 analyzers
LocationBrea, CA, United States

Official product description

ABL800 analyzer with FLEXQ module. Device intended for in vitro testing of samples of whole blood for the parameters pCO2, cK+, cNa+, cCa2+, cCl , cGlu, cLac, cCrea, ctBil, and co-oximetry parameters (ctHb, sO2, and the hemoglobin fractions FO2Hb, FCOHb, FMetHb, FHHb and FHbF) - in vitro testing of samples of expired air for the parameters pO2 and pCO2. It is also used for in vitro testing of pleura samples for the pH parameter

Why the product was recalled

Due to misinterpretation of the barcode by the scanner, when the registration receipt barcode is scanned by the analyzer , a result from a different patient will be printed or displayed on the analyzer.

Distribution information

USA (nationwide) to: AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, WA, WI, WV, WY. Distributed internationally to Canada.

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.

Related recall records

Other records from the same recalling company are shown first when available.

Medical DeviceFeb 13, 2019

Radiometer ABL800 FLEX with Crea recall — FDA Z-0801-2019

From three incidents in Denmark and Sweden, comparison studies on patient blood samples carried out for quality assurance demonstrated that creatinine measurements as determined on…

Class IIFDA: TerminatedRadiometer America Inc.