FDA recall record

Radiometer ABL90 FLEX Analyzer REF 393090 UDI:05700693930909 recall — FDA Z-2320-2019

Software Security; The action is being initiated because of software security vulnerabilities with the firm's analyzer operating system, which may cause…

Class IIclassification
Z-2320-2019official record ID
Aug 28, 2019report date

Recall record at a glance

Official recordZ-2320-2019
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusTerminated
Recalling firmRadiometer America Inc.
Report dateAug 28, 2019
Recall initiation dateJul 10, 2019
Quantity13,042 units
LocationBrea, CA, United States

Official product description

ABL90 FLEX Analyzer REF 393090 UDI:05700693930909

Why the product was recalled

Software Security; The action is being initiated because of software security vulnerabilities with the firm's analyzer operating system, which may cause the device to shutdown or reboot resulting in delayed medical treatment.

Distribution information

US: AB, AL, AR, BC, CA, CO, CT, FL, GA, IA, ID, IL, IN, KY, LA, MA, MD, ME, MI, MN, MO, MS, NC, NE, NH, NJ, NY, OH, OK, OR, PA, SC, TX, VA, VT, WA, WI, WV OUS: Canada, Netherland, China, Czech Republic, Denmark, Germany, Spain, France, Hugary, , indiat Japan, South Korea, United Kingdom, Australia, Poland, South Africa, Singapore, Switzerland, Turkey, Norway, Austria, Italy, Sweden, Finland, Brazil, and Mexico

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: E-Mail.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.

Related recall records

Other records from the same recalling company are shown first when available.

Medical DeviceFeb 13, 2019

Radiometer ABL800 FLEX with Crea recall — FDA Z-0801-2019

From three incidents in Denmark and Sweden, comparison studies on patient blood samples carried out for quality assurance demonstrated that creatinine measurements as determined on…

Class IIFDA: TerminatedRadiometer America Inc.