FDA recall record

Radiometer ABL800 FLEX with Crea recall — FDA Z-0801-2019

From three incidents in Denmark and Sweden, comparison studies on patient blood samples carried out for quality assurance demonstrated that creatinine…

Class IIclassification
Z-0801-2019official record ID
Feb 13, 2019report date

Recall record at a glance

Official recordZ-0801-2019
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusTerminated
Recalling firmRadiometer America Inc.
Report dateFeb 13, 2019
Recall initiation dateJul 2, 2018
Quantity1619
LocationBrea, CA, United States

Official product description

Product: ABL800 FLEX with Crea. The ABL800 with Crea is available in the following configurations: ABL817, ABL827 and ABL837. ABL800 Series Ref. 393-803, 393-804, 393-805, 393-806, 393-820, 393-821, 393-825, 393-826, 393-835, 393-836, 393-844, 393-845.

Why the product was recalled

From three incidents in Denmark and Sweden, comparison studies on patient blood samples carried out for quality assurance demonstrated that creatinine measurements as determined on analyzers could be underestimated with up to 28% as compared with measurements from COBAS. The underestimated creatinine occurs specifically, when measurements on Autocheck 6+ level 1 (also known as S7835 Autocheck module and abbreviated AC6+L1) are less than or equal to 230 micro-meter of a specified insert interval for AC6+L1 with insert limits 210 290 micro-meter. Ultimately, a lower read-out on patient samples is observed Customers are concerned about underestimation of creatinine results in patients, which are to undergo radiocontrast x-ray investigations. If these patients are not healthy, the exposure of radiocontrast agents may induce kidney insufficiency (contrast-induced nephropathy). Hazardous situation: Underestimation of the concentration of creatinine.

Distribution information

US: AZ, CA, CO, CT, FL, GA, IA, ID, IL, IN, KS, MA, MD, ME, MI, MN, MO, NC, NE, NJ, NM, NY, OH, OK, PA, TN, TX, VA, WA, WI, and WV OUS: Canada, Croatia, Finland and Russia

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.