Arthrex Low Profile Screw, 4.5 x 20 mm The Low Profile ScrewTM, 4.5 x 20 mm, AR-8545-20, is recall — FDA Z-1510-2020
Screws provided in the device package are 6 mm longer than intended.
Browse the newest complete recall records currently stored from FDA enforcement reports and NHTSA vehicle campaigns. “Latest” refers to the newest source-dated record in this database, not the timestamp of our last API check.
The newest recall records with useful source details are listed here.
Screws provided in the device package are 6 mm longer than intended.
It has been identified that the rheumatoid factor (RF) calibrator values over recover in comparison to the reference material RF serum. This could result in reporting falsely abnormal…
The items identified for recall are labeled as sterile. However, the manufacturer has found that product packaging for the lots listed above may be damaged, potentially exposing the product…
A capacitor inside the converter of the Velara X-ray generator may fail after a large number of surges in a short period of time. When this occurs, no image acquisition is possible.
A capacitor inside the converter of the Velara X-ray generator may fail after a large number of surges in a short period of time. When this occurs, no image acquisition is possible.
The medical procedure trays contain one or more recalled surgical gowns. The gowns were manufactured at locations that did not maintain proper environmental conditions. There is no…
Philips has received a number of reports of HeartStart MRx Monitor/Defibrillators that have suffered internal damage and were not able to deliver therapy after having been dropped or…
The medical procedure trays contain one or more recalled surgical gowns. The gowns were manufactured at locations that did not maintain proper environmental conditions. There is no…
Infusion Pump Systems may have a delivery inaccuracy of up to 13%, which may result in 1) faster than expected drug delivery when infusing at a very low rate (0.1 mL/h), or 2) slower than…
A capacitor inside the converter of the Velara X-ray generator may fail after a large number of surges in a short period of time. When this occurs, no image acquisition is possible.
Potential diff. in the length of the sheath tube and the sheath/probe. A diff. in length could lead to the base of the sheath/probe not being fully seated in the sheath tube which could…
The Ysio system may process incorrect internal values for the position of its ceiling tube stand. If in this situation the user activates longitudinal or transversal stand movement (by…
Due to high discordant Estradiol results which.potentially affects all patient populations, A falsely elevated estradiol level could lead a clinician to misinterpret a patient as…
The sterile packaging has the potential to be compromised.
Boston Scientific released a new Software Update Application that was downloaded to four (4) distributed Model 3300 programmers. The Software Update Application then directed these…
A capacitor inside the converter of the Velara X-ray generator may fail after a large number of surges in a short period of time. When this occurs, no image acquisition is possible.
Low Density Polyethylene (LDPE) particles may be present on the surface of the implant upon the opening of the inner packaging.
Low Density Polyethylene (LDPE) particles may be present on the surface of the implant upon the opening of the inner packaging.
A software issue could potentially cause the stand and table movements to be blocked.
The medical procedure trays contain one or more recalled surgical gowns. The gowns were manufactured at locations that did not maintain proper environmental conditions. There is no…
The items identified for recall are labeled as sterile. However, the manufacturer has found that product packaging for the lots listed above may be damaged, potentially exposing the product…
Low Density Polyethylene (LDPE) particles may be present on the surface of the implant upon the opening of the inner packaging.
The medical procedure trays contain one or more recalled surgical gowns. The gowns were manufactured at locations that did not maintain proper environmental conditions. There is no…
A capacitor inside the converter of the Velara X-ray generator may fail after a large number of surges in a short period of time. When this occurs, no image acquisition is possible.