FDA recall record

Boston Scientific, LATITUDE Programming System, Model 3300 - Product Usage:The Boston recall — FDA Z-1511-2020

Boston Scientific released a new Software Update Application that was downloaded to four (4) distributed Model 3300 programmers. The Software Update…

Class IIclassification
Z-1511-2020official record ID
Mar 25, 2020report date

Recall record at a glance

Official recordZ-1511-2020
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusTerminated
Recalling firmBoston Scientific Corporation
Report dateMar 25, 2020
Recall initiation dateFeb 12, 2020
Quantity4 devices
LocationSaint Paul, MN, United States

Official product description

Boston Scientific, LATITUDE Programming System, Model 3300 - Product Usage:The Boston Scientific LATITUDE Programming System is intended for use in hospital and clinical environments to communicate with Boston Scientific implantable systems. The software in use controls all communication functions for the Pulse Generator (PG). Detailed software applications instructions can be found in the associated product literature for the PG being interrogated.

Why the product was recalled

Boston Scientific released a new Software Update Application that was downloaded to four (4) distributed Model 3300 programmers. The Software Update Application then directed these programmers to download software applications that have not been approved by the US Food and Drug Administration.

Distribution information

US Nationwide distribution in the states of NC, TN, and MO.

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.

Related recall records

Other records from the same recalling company are shown first when available.