FDA recall record

CME America BodyGuard Infusion Pump System, All models of BG 323, BG 121, BG 545, BG CV545 recall — FDA Z-1474-2020

Infusion Pump Systems may have a delivery inaccuracy of up to 13%, which may result in 1) faster than expected drug delivery when infusing at a very low…

Class Iclassification
Z-1474-2020official record ID
Mar 25, 2020report date

Recall record at a glance

Official recordZ-1474-2020
CategoryMedical Device Recalls
ClassificationClass I
FDA record statusTerminated
Recalling firmCME America, LLC
Report dateMar 25, 2020
Recall initiation dateJan 7, 2020
Quantity30,596 pumps
LocationGolden, CO, United States

Official product description

CME America BodyGuard Infusion Pump System, All models of BG 323, BG 121, BG 545, BG CV545, BG 575, BG CV575, and CMExpress

Why the product was recalled

Infusion Pump Systems may have a delivery inaccuracy of up to 13%, which may result in 1) faster than expected drug delivery when infusing at a very low rate (0.1 mL/h), or 2) slower than expected drug delivery when infusing at high flow rates (greater-than 500 mL/h)

Distribution information

U.S.: HI, OR, AR, MA, PA, OH, WY, CO, NE, NC, WA, TX, UT, MD, CA, KY, FL, NH, TN, NY, SD, MO, IL, WI, MN, RI, VA, DC, SC, NJ, ME, KS, CT, MI, LA, AK, NM, IA, MT, ID, AZ, AL, VT, IN, MS

How to read the classification

FDA Class I indicates a situation in which there is a reasonable probability that use of or exposure to the product will cause serious adverse health consequences or death. The classification shown here comes directly from the FDA enforcement record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.

Related recall records

Other records from the same recalling company are shown first when available.