FDA recall record

Siemens Healthcare Diagnostics IMMULITE 2000 Systems HCG [200 Tests, material number recall — FDA Z-1015-2022

Potential for falsely elevated hCG results due to sample carryover in serum and urine patient samples, as well as quality control samples and adjustors.

Class IIclassification
Z-1015-2022official record ID
May 4, 2022report date

Recall record at a glance

Official recordZ-1015-2022
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusOngoing
Recalling firmSiemens Healthcare Diagnostics, Inc.
Report dateMay 4, 2022
Recall initiation dateFeb 22, 2022
Quantity10,755 boxes (533 US, 10,222 OUS)
LocationTarrytown, NY, United States

Official product description

IMMULITE 2000 Systems HCG [200 Tests, material number 10381206, catalog number L2KCG] and [600 Tests, material number 10381194, catalog number L2KCG6]

Why the product was recalled

Potential for falsely elevated hCG results due to sample carryover in serum and urine patient samples, as well as quality control samples and adjustors.

Distribution information

Worldwide distribution - US Nationwide distribution in the states of AZ, CA, CO, FL, GA, HI, ID, IN, MN, NC, NH, NJ, NM, NV, NY, OH, OR, PA, SC, TX, VA, WI, WY and OUS: 81 countries.

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.

Related recall records

Other records from the same recalling company are shown first when available.