Conformis iTotal CR Knee Replacement System, Model No recall — FDA Z-1627-2020
A limited number of tibial trays were labeled with an incorrect serial number and were delivered to the wrong customer accounts.
Complete FDA and NHTSA recall records dated in 2020, organized for browsing by month, category and recalling company.
Select a month to review the recall records reported during that period.
Records are ordered by the best available source date.
A limited number of tibial trays were labeled with an incorrect serial number and were delivered to the wrong customer accounts.
An image on the associated polybag contains an incorrect rinse time of 30 (thirty seconds). The correct rinse time is 1 (one) minute.
Instructions for Use (IFU) and Surgical Cleaning and Sterilization Protocols updated with revised dry time parameters. This recall of IFU occurred in 2018.
Instructions for Use (IFU) and Surgical Cleaning and Sterilization Protocols updated with revised dry time parameters. This recall of IFU occurred in 2018.
Instructions for Use (IFU) and Surgical Cleaning and Sterilization Protocols updated with revised dry time parameters. This recall of IFU occurred in 2018.
At doses of less than 100 mJ, in custom and auto repeat modes, a software error in the excimer laser phototherapy system causes the device to fail to cease firing as long as the foot pedal…
A limited number of tibial trays were labeled with an incorrect serial number and were delivered to the wrong customer accounts.
The firm has become aware that there is a likelihood that a missing snap ring or spanner nut may result in an insufficient mount force that causes the surgical light to fall. If the…
Software defect in TEG Manager impacts the displayed alert for out of range test results. Due to this defect, reference range values received from TEG 5000 and displayed on the TEG Manager…
During an FDA audit, it was discovered that the product was not properly registered with FDA.
Distributed without an approved 510(k)
The firm has become aware of a potential problem that their Catheters may break/fracture and/or kink/bend if improperly removed from the packaging/dispenser hoop or during catheter use.
Pinion gear may disengage from track allowing lift to slip or jolt.
Instructions for Use (IFU) and Surgical Cleaning and Sterilization Protocols updated with revised dry time parameters. This recall of IFU occurred in 2018.
Instructions for Use (IFU) and Surgical Cleaning and Sterilization Protocols updated with revised dry time parameters. This recall of IFU occurred in 2018.
Package is intended to contain one (1) baseplate and one (1) taper adapter; however, devices may be packaged with two (2) baseplates and no (0) adapters.
The firm has become aware of a potential problem with their venous and arterial magnetic catheters may be at risk of experiencing a magnetic deficiency, which may result in the magnets…
The firm has become aware of a potential problem that their Catheters may break/fracture and/or kink/bend if improperly removed from the packaging/dispenser hoop or during catheter use.
During an FDA audit, it was discovered that the product was not properly registered with FDA.
The OBM system functional test (impedance and noise) was not carried out on the affected items prior to release to the customer. The manufacturer cannot attest to the system functionality…
Potential for the NAVIO Soft Tissue Protector to become stuck or bound to the bone pin intraoperatively when inserting the NAVIO bone pins.
Instructions for Use (IFU) and Surgical Cleaning and Sterilization Protocols updated with revised dry time parameters. This recall of IFU occurred in 2018.
Automated External Defibrillators may not detect a patient connection due to mechanical wear-through of the contact plating on the therapy connector. The wear-through could result in the…
The firm has become aware that there is a potential that the glass cover at the distal end of the camera/led lighting on the video laryngoscope could become loose during reprocessing or…
This archive groups records by their reported or recall date. It does not imply that every recall was initiated, announced and resolved during the same calendar year. Open the individual record for the exact dates and source details supplied by FDA or NHTSA.