Abbott Laboratories ARCHITECT BNP Controls Ptoduct Usage: The ARCHITECT BNP Controls are for recall — FDA Z-2079-2019
Presence of a time dependent, stability drift in patient and control results returned from ARCHITECT BNP testing.
Complete FDA and NHTSA recall records dated in July 2019.
These records include useful source details for the selected month.
Presence of a time dependent, stability drift in patient and control results returned from ARCHITECT BNP testing.
One bolt on the V motor was not fully tightened which caused a drift of .13 degrees in alignment resulting in reduced dose output.
Manufacturing defect was found for the cartridge component, which could compromise the sterility barrier of the product
A tolerance issue in the power supply of the generator control above a specific value may cause generator A100 failure. As a result, the X-ray tube will not be supplied with the required…
Certain NAUTILUS DELTA Port Kit product code / lot number combinations may be at risk of experiencing a lack of or a loss of the intravascular ECG signal during port placement procedures…
A tolerance issue in the power supply of the generator control above a specific value may cause generator A100 failure. As a result, the X-ray tube will not be supplied with the required…
Mislabeled - The product's labeling contains an incorrect size for the catheter included in the kit. The kit contains a 21 gauge catheter, however the labeling states a 12 gauge catheter is…
A tolerance issue in the power supply of the generator control above a specific value may cause generator A100 failure. As a result, the X-ray tube will not be supplied with the required…
Administration sets leaked at the filter.
Presence of a time dependent, stability drift in patient and control results returned from ARCHITECT BNP testing.
An update to the Reagents and Specimen Collection and Handling/Preparation for Analysis sections of the IFU was required due to different issues with the various products.
Two issues: 1) The MAC VU360 system may intermittently display an incorrect patient ID or visit number on the screen after scanning the patient s barcode. This may result in the ECG report…
A tolerance issue in the power supply of the generator control above a specific value may cause generator A100 failure. As a result, the X-ray tube will not be supplied with the required…
A tolerance issue in the power supply of the generator control above a specific value may cause generator A100 failure. As a result, the X-ray tube will not be supplied with the required…
The syringes have scale markings that were printed incorrectly. The scale is skewed to varying degrees, resulting in missing and/or partially printed scale numbers and scale lines.
There is a potential for a loss of package integrity that may compromise the sterility of the product.
Potential for prior thumbnails to not display in reverse chronological order and images may not scroll in reverse chronological order when mammography stacked scrolling is enabled. Merge…
There is a potential for a loss of package integrity that may compromise the sterility of the product.
A tolerance issue in the power supply of the generator control above a specific value may cause generator A100 failure. As a result, the X-ray tube will not be supplied with the required…
There is a potential for a loss of package integrity that may compromise the sterility of the product.
Products were not held at an appropriate temperature due to a mechanical malfunction during transportation.
Products were not held at an appropriate temperature due to a mechanical malfunction during transportation.
Products were not held at an appropriate temperature due to a mechanical malfunction during transportation.
Products were not held at an appropriate temperature due to a mechanical malfunction during transportation.
Use this page to compare the recalls reported during the month, then open an individual record for the exact reason, classification, company, remedy or distribution information available from the official source. Monthly counts are descriptive, not a safety rating.