Baxter Healthcare XScribe Cardiac Stress Testing System, version 6, with below product names recall — FDA Z-0706-2024

Potential distortion identified in electrocardiogram (ECG) readings when the Source Consistency Filter (SCF) is enabled.

Class IIclassification
Z-0706-2024official record ID
Jan 17, 2024report date

Recall record at a glance

Official recordZ-0706-2024
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusOngoing
Recalling firmBaxter Healthcare Corporation
Report dateJan 17, 2024
Recall initiation dateDec 8, 2023
Quantity2985 units
LocationDeerfield, IL, United States

Official product description

XScribe Cardiac Stress Testing System, version 6, with below product names and codes: a) XS6/RS6 REVIEW WKSTN KIT, XR-6AC-BXXXX; b) XS6/RS6 REVIEW WKSTN KIT, XR-6AC-DXXXX; c) XS6/RS6 REVIEW WKSTN KIT, XR-6AC-XXXXX; d) XS6 WAM TTL KIT IEC XML BRAZIL, XSCRIBE-5AX-XXBRZ; e) XS6 WAM TTL SYS Z200 AHA XML, XSCRIBE-6AA-AAAAA; f) XS6 WAM TTL SYS Z200 BCRT IEC, XSCRIBE-6AA-ADBXX; g) XS6 WAM TTL SYS Z200 PCRT AHA XML, XSCRIBE-6AA-AFAAA; h) XS6 WAM TTL SYS Z200 PCRT AHA SVR, XSCRIBE-6AA-AFAAC; i) XS6 WAM TTL SYS Z200 PCRT AHA XML, XSCRIBE-6AA-AFAXA; j) XS6 WAM TTL SYS Z200 PCRT AHA DICOM, XSCRIBE-6AA-AFAXB; k) XS6 WAM TTL SYS Z200 BCRT AHA DICOM, XSCRIBE-6AA-BCAAB; l) XS6 WAM TTL SYS AHA XML, XSCRIBE-6AA-BXAAA; m) XS6 WAM TTL SYS AHA DICOM, XSCRIBE-6AA-BXAAB; n) XS6 WAM TTL SYS AHA, XSCRIBE-6AA-BXAXX; o) XS6 WAM TTL SYS Z200 BCRT IEC XML, XSCRIBE-6AA-CDBAA; p) XS6 WAM TTL SYS Z200 BCRT IEC DICOM, XSCRIBE-6AA-CDBAB; q) XS6 WAM TTL SYS Z200 BCRT IEC SVR, XSCRIBE-6AA-CDBAC; r) XS6 WAM TTL SYS Z200 BCRT IEC, XSCRIBE-6AA-CDBAX; s) XS6 WAM TTL SYS BCRT AHA DICOM, XSCRIBE-6AA-CEAAB; t) XS6 WAM TTL SYS BCRT IEC DICOM, XSCRIBE-6AA-CEBAB; u) XS6 WAM TTLSYS Z200 BCRT IEC DICOM, XSCRIBE-6AA-DDBAB; v) XS6 WAM TTL SYS Z200 PCRT AHA XML, XSCRIBE-6AA-DGAAA; w) XS6 WAM TTL SYS Z200 PCRT IEC DICOM, XSCRIBE-6AA-DGBAB; x) XS6 WAM TTL SYS Z200 PCRT IEC, XSCRIBE-6AA-DGBAX; y) XS6 WAM TTL SYS Z200 PCRT AHA XML, XSCRIBE-6AA-XFAAA; z) XS6 WAM TTL SYS Z200 PCRT AHA DICOM, XSCRIBE-6AA-XFAAB; aa) XS6 WAM TTL SYS Z200 PCRT AHA XML, XSCRIBE-6AA-XFAXA; bb) XS6 WAM TTL SYS Z200 PCRT IEC XML, XSCRIBE-6AA-XFBXA; cc) XS6 WAM TTL SYS BCRT AHA, XSCRIBE-6AB-BEAXX; dd) XS6 WAM TTL SYS AHA XML, XSCRIBE-6AB-BXAXA; ee) XS6 WAM TTL SYS Z200 BCRT AHA XML, XSCRIBE-6AB-CDAAA; ff) XS6 WAM TTL SYS Z200 BCRT AHA DICOM, XSCRIBE-6AB-CDAAB; gg) XS6 WAM TTLSYS Z200 BCRT IEC DICOM, XSCRIBE-6AB-CDBAA; hh) XS6 WAM TTL SYS Z200 BCRT IEC DICOM, XSCRIBE-6AB-CDBAB; ii) XS6 WAM TTL SYS Z200 BCRT IEC SVR, XSCRIBE-6AB-CDBAC; jj) XS6 WAM TTL SYS BCRT AHA XML, XSCRIBE-6AB-CEAAA; kk) XS6 WAM TTL SYS BCRT IEC, XSCRIBE-6AB-CEBAX; ll) XS6 WAM TTL SYS Z200 BCRT IEC DICOM, XSCRIBE-6AB-DDBAB; mm) XS6 WAM TTL SYS Z200 BCRT IEC, XSCRIBE-6AB-DDBAX; nn) XS6 WAM TTL SYS BCRT IEC XML, XSCRIBE-6AB-DEBAA; oo) XS6 WAM TTL SYS BCRT IEC DICOM, XSCRIBE-6AB-DEBAB; pp) XS6 WAM TTL SYS BCRT IEC SVR, XSCRIBE-6AB-DEBAC; qq) XS6 WAM TTL SYS AHA XML, XSCRIBE-6AB-DXAAA; rr) XS6 WAM TTL SYS AHA XML, XSCRIBE-6AB-DXAXA; ss) XS6 WAM TTLSYS Z200 BCRT AHA XML, XSCRIBE-6AB-XCAAA; tt) XS6 WAM TTLKIT Z200 AHA DICOM, XSCRIBE-6AC-AAAAB; uu) XS6 WAM TTL KIT Z200 AHA, XSCRIBE-6AC-AAAAX; vv) XS6 WAM TTL KIT Z200 AHA, XSCRIBE-6AC-AAAXX; ww) XS6 WAM TTL KIT Z200 BCRT AHA, XSCRIBE-6AC-ACAAX; xx) XS6 WAM TTL KIT Z200 BCRT AHA, XSCRIBE-6AC-ACAXX; yy) XS6 WAM TTL KIT BCRT AHA XML, XSCRIBE-6AC-AEAAA; zz) XS6 WAM TTL KIT BCRT AHA, XSCRIBE-6AC-AEAXX; aaa) XS6 WAM TTL KIT AHA XML, XSCRIBE-6AC-AXAAA; bbb) XS6 WAM TTL KIT AHA XML, XSCRIBE-6AC-AXAXA; ccc) XS6 WAM TTL KIT AHA, XSCRIBE-6AC-AXAXX; ddd) XS6 WAM TTL KIT Z200 BCRT AHA DICOM, XSCRIBE-6AC-BDAAB; eee) XS6 WAM TTL KIT BCRT AHA SVR, XSCRIBE-6AC-BEAAC; fff) XS6 WAM TTL KIT BCRT AHA, XSCRIBE-6AC-BEAXX; ggg) XS6 WAM TTL KIT AHA XML, XSCRIBE-6AC-BXAXA; hhh) XS6 WAM TTL KIT AHA, XSCRIBE-6AC-BXAXX; iii) XS6 WAM TTL KIT Z200 BCRT IEC XML, XSCRIBE-6AC-CDBAA; jjj) XS6 WAM TTL KIT Z200 BCRT IEC DICOM, XSCRIBE-6AC-CDBAB; kkk) XS6 WAM TTL KIT BCRT IEC, XSCRIBE-6AC-CEBAX; lll) XS6 WAM TTL KIT IEC XML, XSCRIBE-6AC-CXBAA; mmm) XS6 WAM TTL KIT IEC DICOM, XSCRIBE-6AC-CXBAB; nnn) XS6 WAM TTL KIT Z200 IEC XML, XSCRIBE-6AC-DBBXA; ooo) XS6 WAM TTL KIT Z200 BCRT IEC XML, XSCRIBE-6AC-DCBXA; ppp) XS6 WAM TTL KIT Z200 BCRT IEC XML, XSCRIBE-6AC-DDBAA; qqq) XS6 WAM TTL KIT Z200 BCRT IEC DICOM, XSCRIBE-6AC-DDBAB; rrr) XS6 WAM TTL KIT Z200 BCRT IEC DICOM, XSCRIBE-6AC-DDBXB; sss) XS6 WAM TTL KIT BCRT IEC XML, XSCRIBE-6AC-DEBAA; ttt) XS6 WAM TTL KIT BCRT IEC

Why the product was recalled

Potential distortion identified in electrocardiogram (ECG) readings when the Source Consistency Filter (SCF) is enabled.

Distribution information

United States (including Puerto Rico), Canada, Australia, Bangladesh, Brunei, Chia, Hong Kong, India, Indonesia, Korea, Malaysia, New Zealand, Philippines, Singapore, Sri Lanka, Thailand, Algeria, Angola, Austria, Bahrain, Bulgaria, Croatia, Cyprus, Czech Republic, Denmark, Egypt, Estonia, Finland, France, Georgia, Germany, Ghana, Greece, Hungary, Iceland, Iran, Iraq, Ireland, Israel, Italy, Jordan, Kazakhstan, Kenya, Kuwait, Latvia, Lebanon, Libya, Morocco, Namibia, Netherlands, Norway, Oman, Pakistan, Poland, Portugal, Qatar, Romania, Russia, Saudi Arabia, Serbia, South Africa, Spain, Switzerland, Tunisia, Turkey, United Arab Emirates, United Kingdom, Argentina, Bahamas, Bolivia, Brazil, Chile, Colombia, Costa Rica, Dominican Republic, Ecuador, Guatemala, Honduras, Mexico, Panama, Paraguay, Peru, and Trinidad and Tobago.

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.