FDA recall record

Baxter Healthcare Hillrom VOLARA System P.CIRCUIT KIT AC recall — FDA Z-2550-2026

There have been reports of air and medication leakage from the nebulizer cup during therapy of Volara system patient circuits. The leakage can lead to an…

Class Iclassification
Z-2550-2026official record ID
Jul 8, 2026report date

Recall record at a glance

Official recordZ-2550-2026
CategoryMedical Device Recalls
ClassificationClass I
FDA record statusOngoing
Recalling firmBaxter Healthcare Corporation
Report dateJul 8, 2026
Recall initiation dateMay 21, 2026
Quantity16187 units
LocationDeerfield, IL, United States

Official product description

Hillrom VOLARA System P.CIRCUIT KIT AC (Single Patient Use Circuit), Model/Catalog Number REF M08473; Oscillation Lung Expansion Therapy, positive pressure breathing device

Why the product was recalled

There have been reports of air and medication leakage from the nebulizer cup during therapy of Volara system patient circuits. The leakage can lead to an oxygen level drop in the patient and ineffective nebulization, which may affect the delivery of the prescribed treatment. The issue has been associated with improper locking of the nebulizer cup after adding medication during user setup.

Distribution information

Worldwide distribution - US Nationwide and the country of Canada.

How to read the classification

FDA Class I indicates a situation in which there is a reasonable probability that use of or exposure to the product will cause serious adverse health consequences or death. The classification shown here comes directly from the FDA enforcement record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.

Related recall records

Other records from the same recalling company are shown first when available.

Medical DeviceJun 10, 2026

Exacta Mix Vented Micro-Volume Inlet recall — FDA Z-2262-2026

Affected lots may contain incorrect 0.8-micron filters. Use of a 0.8-micron filter in place of the validated 3-micron filter may lead to reduced flow efficiency, incomplete or slower…

Class IIFDA: OngoingBaxter Healthcare Corporation