FDA recall record

Synthes USA HQ Variable Angle Locking Compression Plate/VA-LCP Dorsal Distal Radius Plate recall — FDA Z-0956-2013

It was discovered that one of the VA-LCP Dorsal Distal Radius Plates was incorrectly etched as Part Number 02.115.431 (110 degree) instead of the correct…

Class IIclassification
Z-0956-2013official record ID
Mar 27, 2013report date

Recall record at a glance

Official recordZ-0956-2013
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusTerminated
Recalling firmSynthes USA HQ, Inc.
Report dateMar 27, 2013
Recall initiation dateJan 11, 2013
Quantity4
LocationWest Chester, PA, United States

Official product description

Variable Angle Locking Compression Plate/VA-LCP Dorsal Distal Radius Plate, Part Number 02.115.431. Indicated for: Dorsally displaced fractures, Open joint reconstruction, and Corrective osteotomies after distal radius malunion.

Why the product was recalled

It was discovered that one of the VA-LCP Dorsal Distal Radius Plates was incorrectly etched as Part Number 02.115.431 (110 degree) instead of the correct Part Number 02.115.231 (90 degree).

Distribution information

Distributed in the states of IN, OH, PA, and TX.

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.

Related recall records

Other records from the same recalling company are shown first when available.