Recall record at a glance
| Official record | Z-0473-2014 |
|---|---|
| Category | Medical Device Recalls |
| Classification | Class II |
| FDA record status | Terminated |
| Recalling firm | Synthes USA HQ, Inc. |
| Report date | Dec 18, 2013 |
| Recall initiation date | Oct 9, 2013 |
| Quantity | 2912 |
| Location | West Chester, PA, United States |
Official product description
Synthes TomoFix Plates The Synthes J5606-C TomoFix Osteotomy System is used for open and closed wedge osteotomies of the medial proximal tibia, lateral proximal tibia, medial and lateral distal femur, treatment of bone and joint deformities, fractures, and malalignment caused by injury or disease such as osteoarthritis.
Why the product was recalled
A labeling correction was initiated related to the Surgical Technique Guide J5606-C "TomoFix Osteotomy System" due to the potential for a surgical delay if the surgeon inserts the superior screws at a lesser angle (4 or 6 degrees) and is not aware that those angles are not present.
Distribution information
Worldwide Distribution - USA (nationwide) Canada and Mexico
How to read the classification
FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.
Additional record details
Recall type: Voluntary: Firm initiated. Initial notification: Letter.
What to do next
If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.
About this record
Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.
FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.