FDA recall record

DRG International Total Thyroxine (T4) ELISA recall — FDA Z-0310-2022

Firm has identified two lots of product that do not meet product shelf life, resulting in failed Controls. The clinical impact may be a delay in diagnosis…

Class IIclassification
Z-0310-2022official record ID
Dec 8, 2021report date

Recall record at a glance

Official recordZ-0310-2022
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusOngoing
Recalling firmDRG International, Inc.
Report dateDec 8, 2021
Recall initiation dateOct 18, 2021
Quantity314 total (270 US)
LocationSpringfield, NJ, United States

Official product description

Total Thyroxine (T4) ELISA; REF EIA-1781

Why the product was recalled

Firm has identified two lots of product that do not meet product shelf life, resulting in failed Controls. The clinical impact may be a delay in diagnosis and a consumption of patient samples.

Distribution information

Domestic: NY; International: TZ, VN, BD, CO, TT, YE, BO, DE, SV, GE, PY, TG, RU/UA, NI, DO, NG, EG & CM

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: E-Mail.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.

Related recall records

Other records from the same recalling company are shown first when available.

Medical DeviceAug 13, 2025

T4 Total ELISA recall — FDA Z-2265-2025

An incoming complaint reported that the expiration date on the kit box label and Certificate of Analysis (CoA) exceeded the shelf life of the standards and control solutions in the kit by 2…

Class IIIFDA: OngoingDRG International, Inc.