FDA recall record

DRG International 17-OH Progesterone ELISA 96 well kit- An enzyme immunoassay for the recall — FDA Z-0258-2025

The High and Low Control acceptance ranges in the Certificate of Analysis (CofA) did not match the acceptance range information on the High and Low…

Class IIIclassification
Z-0258-2025official record ID
Nov 6, 2024report date

Recall record at a glance

Official recordZ-0258-2025
CategoryMedical Device Recalls
ClassificationClass III
FDA record statusOngoing
Recalling firmDRG International, Inc.
Report dateNov 6, 2024
Recall initiation dateSep 13, 2024
Quantity504 units
LocationSpringfield, NJ, United States

Official product description

17-OH Progesterone ELISA 96 well kit- An enzyme immunoassay for the quantitative in vitro diagnostic measurement of 17-OH Progesterone (17-OHP) in serum or plasma Model/Catalog Number: EIA-1292

Why the product was recalled

The High and Low Control acceptance ranges in the Certificate of Analysis (CofA) did not match the acceptance range information on the High and Low Control Component labels. Due to the defect, the device failed to provide results, which may result in a delay in obtaining test results and a remote chance of an adverse health risk.

Distribution information

AZ, CA, NH Foreign: Austria Belgium Georgia Germany Greece India Irak Italy Netherlands Poland Romania Russia Singapore Spain Tunesia Vietnam

How to read the classification

FDA Class III generally covers situations in which use of or exposure to the product is not likely to cause adverse health consequences. The classification displayed here is taken from the FDA enforcement record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: E-Mail.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.