FDA recall record

Intuitive Surgical Thoracic Grasper, 5 MM, for use with da Vinci S (IS2000) and SI (IS3000) recall — FDA Z-1812-2015

The coating of the 5 mm Thoracic Grasper main tube may experience damage with the possibility of fragments of coating material dislodging and potentially…

Class IIclassification
Z-1812-2015official record ID
Jun 24, 2015report date

Recall record at a glance

Official recordZ-1812-2015
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusTerminated
Recalling firmIntuitive Surgical, Inc.
Report dateJun 24, 2015
Recall initiation dateMay 21, 2015
Quantity2,436 units
LocationSunnyvale, CA, United States

Official product description

Thoracic Grasper, 5 MM, for use with da Vinci S (IS2000) and SI (IS3000) Surgical Systems. Intended for endoscopic manipulation of tissue, including grasping.

Why the product was recalled

The coating of the 5 mm Thoracic Grasper main tube may experience damage with the possibility of fragments of coating material dislodging and potentially falling into patient.

Distribution information

Worldwide Distribution -- US, Belgium, Brazil, Canada, Chile, Colombia,, Denmark, France, Germany, India, Netherlands, Panama, Singapore, South Korea, Switzerland, Taiwan, Turkey, and United Kingdom.

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.