FDA recall record

Intuitive Surgical Da Vinci ASSY, PSS, SP1098, Part Number: 380601 recall — FDA Z-2521-2026

Operative components of surgical systems potentially distributed with E-brake retainer pin, so in event of a structural column failure, the E-brake pin…

Class IIclassification
Z-2521-2026official record ID
Jul 1, 2026report date

Recall record at a glance

Official recordZ-2521-2026
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusOngoing
Recalling firmIntuitive Surgical, Inc.
Report dateJul 1, 2026
Recall initiation dateApr 22, 2026
Quantity454
LocationSunnyvale, CA, United States

Official product description

Da Vinci ASSY, PSS, SP1098, Part Number: 380601

Why the product was recalled

Operative components of surgical systems potentially distributed with E-brake retainer pin, so in event of a structural column failure, the E-brake pin would prevent E-brake engagement and may result in uncontrolled motion. In addition, if it moves/backs out it may cause restricted column range of motion and/or uncontrolled motion, which may cause patient tissue injury/crushing injury/bleeding.

Distribution information

Worldwide - US Nationwide distribution including in the states of CA, GA, OH, PA, NY, MN, TN, OR, IL, MD, NJ, AL, NV, TX, FL, LA, CT, NC, AZ, IA, MO, WA, MI, WV, WI, KY, KS, VA, OK, MA, IN, CO, SC, SD, MS, NH, UT, ME and the countries of Belgium, Denmark, Sweden, France, Japan, Germany, Israel, Switzerland, South Korea, Czech Republic, Austria, United Kingdom, Italy, Spain, Ireland, Saudi Arabia, Greece, Taiwan, Australia, China, Turkey, Poland, United Arab Emirates, Hong Kong, Romania.

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.

Related recall records

Other records from the same recalling company are shown first when available.

Medical DeviceJun 10, 2026

Da Vinci 5 ASSY,DV5 TOWER,IS5000 recall — FDA Z-2310-2026

Due to unintentionally activated near infrared handheld camera when attempting to turn off illumination during normal clinical workflow transitions, particularly when switching from the…

Class IIFDA: OngoingIntuitive Surgical, Inc.