Recall record at a glance
| Official record | Z-0008-2017 |
|---|---|
| Category | Medical Device Recalls |
| Classification | Class II |
| FDA record status | Terminated |
| Recalling firm | Microgenics Corporation |
| Report date | Oct 12, 2016 |
| Recall initiation date | Aug 15, 2016 |
| Quantity | 163 kits in US, 669 kits worldwide - toatal 831. |
| Location | Fremont, CA, United States |
Official product description
Thermo Scientific QMS Everolimus Assay; Model number: 0380000 (US Distribution); 0373852 (Foreign Distribution) Product Usage: The QMS Everolimus Assay is intended for the quantitative determination of Everolimus, in human whole blood on automated clinical chemistry analyzers. The results obtained are used as an aid in the management of kidney and liver transplant patients receiving Everolimus therapy. This in vitro diagnostic device is intended for clinical laboratory use only. The QMS Everolimus assay is intended to be used as an aid in the management of patients receiving everolimus therapy for those organ transplant procedures indicated in the chart for each specific country.
Why the product was recalled
A reagent stability issue (still under investigation) causes less separation in the calibration curve, which affects results across the entire range. For controls, maximum bias of +/- 35% for low levels, +/- 33% for mid-range levels, and +/- 47% for high levels is observed. Bias in patient samples ranges from -15.7% to +5.2% with an average bias of -5.6% when comparing the QMS everolimus assay to an everolimus LC-MS reference method. Patient samples across the measuring range sub-therapeutic, therapeutic, and supra-therapeutic could be affected.
Distribution information
Worldwide Distribution - US Nationwide in the states of KY, CA, NY, DC, NC, TX, IL, OH, IL, FL, IK, IA, MS and countries of Mexico, Taiwan, Guatemala, Malaysia, Argentina, Dominican Republic, Philippines, Germany, Australia.
How to read the classification
FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.
Additional record details
Recall type: Voluntary: Firm initiated. Initial notification: Letter.
What to do next
If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.
About this record
Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.
FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.